ALMA Memo #461 The Amplitude Calibration System revisited S.Guilloteau (IRAM / ESO) A.Bacmann (ESO) 16 Sep 2003 Keywords: Amplitude Calibration, Receiver Saturation Different strategies of use of a semi-transparent device combined with two temperature loads for calibration are explored. It is shown that a system with an ambient load, a hot (100 Celsius or so) load, and a semi-transparent grid with 50 % transmission allows to reach a 1 % accuracy on the receiver gain even in presence of saturation. The best precision requires a device which offers the 5 major combinations: Sky, Sky+Ambient, Sky+Hot, Ambient, Hot. If saturation is negligible, the system can be used as a standard dual load calibration device, with better accuracy.